Glove Box Regeneration Actions

From The Third Age: Wiki
Jump to: navigation, search

one) Regeneration fuel: a combination of hydrogen and inert gas (hydrogen accounts for 5% -10%)

two) Amount: 40L gasoline cylinder 10Mpa

three) Regenerative fuel pressure: .05Mpa (the gasoline consumption for every fifty percent hour in the state of .05Mpa is that the cylinder stress drops by one.5Mpa. If it exceeds 1.5Mpa, the stress of the pressure regulator’s timetable will be reduced, if it is considerably less than 1.5Mpa Boost the tension of the pressure regulator program)

4) Course of action: 3 hrs heating-three hours heating and glove box manufacturers air flow-3 several hours vacuuming-one hour pumping-ten hours cooling

5) Ways:

① Hook up the regeneration gas (verify the pipeline for leaks immediately after link: following connecting, open the primary valve of the pressure lowering valve to the most, open the auxiliary meter to .1Mpa, transform off the primary valve, wait around 15 minutes, if the stress of the auxiliary meter remains at .1 Mpa remains unchanged, indicating no air leakage

>②Set the strain of the auxiliary gauge of the strain lowering valve to .1Mpa (the strain in the course of air flow is about .06-.08MPa

>③ Close the cycle and analyz

r>④Turn on the vacuum pu

r>⑤ Commence regenerat

br>⑥ The new system finishes at 20h (clean ahead of cleaning, drinking water and oxygen index <200p


6) Note: Do not cut off the power and gas during the regeneration process (especially the first 6 hours)! Some pressure reducing valves are not allowed, and the pressure needs to be adjusted appropriately to ensure continuous ventilation for 3 ho


7) Observation: whether the regeneration gas flowmeter shows about 20L / min, otherwise adjust the pressure of the pressure reducing valve.